Ted-med Equipments Private Limited
Ted-med Equipments Private Limited
Panipat, Haryana
GST No. 06AAICT4619D1ZG
TrustSEAL Verified
Call 08047642515 64% Response Rate
SEND EMAIL

Drager Vaporizer

Prominent & Leading Wholesaler from Panipat, we offer drager halothane vaporizer, drager desflurane vaporizer and drager sevo flurane vaporizer.

Drager Halothane Vaporizer

Drager Halothane Vaporizer
  • Drager Halothane Vaporizer
  • Drager Halothane Vaporizer
  • Drager Halothane Vaporizer
  • Drager Halothane Vaporizer
  • Drager Halothane Vaporizer
  • Drager Halothane Vaporizer
Get Best Quote
Approx. Price: Rs 18,000 / PieceGet Latest Price
Product Brochure

Product Details:

Minimum Order Quantity1 Piece
Agent TypeHalothane
Mounting TypeSelectatec
Filling TypeKey fill
Output Range2 - 6 %
MaterialStainless Steel
Model Name/NumberAnesthesia Vaporizer
BrandRefurb Imported
Inlet Port23 mm
Temperature CompYes
Country of OriginImported
Usage/ApplicationICU Use
Automation GradeManual

The Dräger Halothane Vaporizer (most commonly the Vapor 19.3 or Vapor 2000 series) is a precision-engineered, agent-specific device used in anesthesia machines to vaporize liquid halothane and add it to the fresh gas flow (oxygen/air/nitrous oxide) delivered to a patient. Here is a detailed breakdown of its features, design, and function:1. Key Characteristics and Design
  • Agent-Specific: The Dräger vaporizer is exclusively calibrated for halothane. It features a color-coded (usually red for halothane) key-indexed filling system to prevent errors.
  • Variable Bypass Principle: The vaporizer splits the incoming fresh gas flow into two streams: one enters the "vaporizing chamber" (to pick up anesthetic) and the other passes through the "bypass chamber" (to skip it). These streams recombine before leaving the vaporizer.
  • Temperature Compensation: It features an internal temperature-sensitive element (bimetallic strip) that automatically adjusts the ratio of gas flow. When temperatures drop, less gas passes through the vaporizing chamber, maintaining a stable, consistent concentration.
  • High Thermal Capacity: The vaporizers are constructed with large masses of metal (e.g., aluminum) to act as a heat sink. This counteracts the cooling effect that occurs during evaporation (latent heat of vaporization).
  • Mounting System: Compatible with Dräger Interlock or Selectatec systems, allowing quick, secure, and safe attachment to the anesthesia machine, often featuring a locking mechanism to prevent multiple vaporizers from being turned on simultaneously. 
2. Operational Details
  • Concentration Range: Typically calibrated to deliver halothane concentrations between 0.2% and 4–5% by volume.
  • Flow Range: Accurately maintains vapor concentration over a wide fresh gas flow range (typically 250 mL/min to 15 L/min).
  • Filling Mechanism: Designed for key-indexed fillers, which allow filling without exposing the operator to liquid anesthetic and prevent overfilling.
  • Transport Mode (Vapor 2000): The Vapor 2000 series allows for transport in a tilted or upside-down position without draining, provided it is turned to the "T" position. 
Request
Callback


Additional Information:

  • Item Code: Drager
  • Delivery Time: 5
Yes! I am Interested

Drager Desflurane Vaporizer

Drager Desflurane Vaporizer
  • Drager Desflurane Vaporizer
  • Drager Desflurane Vaporizer
  • Drager Desflurane Vaporizer
  • Drager Desflurane Vaporizer
  • Drager Desflurane Vaporizer
  • Drager Desflurane Vaporizer
Get Best Quote
Approx. Price: Rs 35,000 / PieceGet Latest Price
Product Brochure

Product Details:

Minimum Order Quantity1 Piece
Agent TypeDesflurane
Mounting TypeSelectatec
Filling TypeKey fill
Positioning TypeTrolley
Output Range0–18%
Calibration GasO2/AIR mixture
Temperature CompYes
Gas Sourceyes
Usage/ApplicationAnesthesia vaporizer
Monitor Shelfyes
Inlet PortDrager port
MaterialAluminium alloy
Country of OriginIndia

Instructions for use D-Vapor/D-Vapor 3000  
WARNING Anesthetic agent vaporizer To properly use this medical device, read and comply with these instructions for use.
Typographical conventions Use of terms Dräger uses the term "accessory" not only for accessories in the sense of IEC 60601-1, but also for consumable parts, removable parts, and attached parts. These instructions for use apply to D-Vapor and D-Vapor 3000 anesthetic vaporizers. If information only applies to one model of anesthetic vaporizer, this is noted in the heading or in the corresponding part of the text. Trademarks Safety Information Definitions 1 Consecutive numbers indicate steps of action, with the numbering restarting with "1" for each new sequence of actions.  Bullet points indicate individual actions or different options for action. – Dashes indicate the listing of data, options, or objects. (A) Letters in parentheses refer to elements in the related illustration. A Letters in illustrations denote elements referred to in the text.
Request
Callback


Additional Information:

  • Delivery Time: 5 days
Yes! I am Interested

Drager Sevo Flurane Vaporizer

Drager Sevo Flurane Vaporizer
  • Drager Sevo Flurane Vaporizer
  • Drager Sevo Flurane Vaporizer
  • Drager Sevo Flurane Vaporizer
  • Drager Sevo Flurane Vaporizer
  • Drager Sevo Flurane Vaporizer
Get Best Quote
Approx. Price: Rs 35,000 / PieceGet Latest Price
Product Brochure

Product Details:

Minimum Order Quantity1 Piece
Battery Backup Time45 min
Inspiration time (Tinsp)1.2
Mains power connectionyes
Country of OriginMade in India

The Dräger Vapor 2000 and Vapor 3000 series are high-precision, variable-bypass vaporizers designed specifically for the controlled vaporization of liquid sevoflurane, an essential agent for general anesthesia. They are widely used in modern operating rooms to accurately add anesthetic vapor to fresh gas flow.Key Features & Characteristics
  • Agent-Specific: Engineered specifically for sevoflurane to ensure accurate delivery, preventing the mix-up of volatile agents.
  • Temperature & Flow Compensation: These vaporizers automatically adjust to variations in temperature (operating range 10°C to 40°C) and fresh gas flow (250 mL/min to 15 L/min), maintaining consistent output.
  • Transport Position ("T"): The Vapor 2000/3000 series can be tilted or turned upside down during transport without leaking, enabling immediate use upon reconnection.
  • Large Capacity: The unit can hold the entire content of a standard 250 mL anaesthetic agent bottle (up to 300 mL), allowing for less frequent refills.
  • Filling System Options: Available with Key-Indexed Filling (KIF) or filling spouts, ensuring safe, agent-specific filling to prevent incorrect agent administration.
  • Maintenance-Free: The Vapor 3000 series requires no routine preventive maintenance or recalibration over its lifespan.
  • Compatibility: Compatible with Dräger Interlock 2 and Selectatec mounting systems for easy integration with anesthetic workstations. 
Request
Callback


Additional Information:

  • Item Code: 4582
  • Production Capacity: VERY GOOD
  • Delivery Time: 3-4 DAYS
  • Packaging Details: Dräger vaporisers have been the benchmark for quality for over 50 years. Quality trusted by doctors and nurses around the world: to date, over 400, 000 Vapor units have been sold to hospitals around the world. Simply more volume: 300
Yes! I am Interested
X

Product Videos

Explore More Products

View All Products
Tell Us What Are You Looking For ?




Reach Us
Mr. Mosin Siddqui (CEO)
Ted-med Equipments Private Limited
Ward-17 Near Unique Public School New Mukhija Colony
Panipat - 132103, Haryana, India
Get Directions

Call Us


Send E-mail